DI.MPU 01

“TRANSIT TIME” FLOW METER

THE DSP TECHNOLOGY (DIGITAL SIGNAL PROCESSING) ENSURE A LOW SENSIBILITY OF THE INSTRUMENT AGAINST POTENTIAL TRANSIENT FACTORS. THE SIZE OF THE PIPE MAY VARY 20 ÷ 4000 MM (USING DIFFERENT TRANSDUCERS). SINCE THE TRANSDUCERS ARE APPLIED EXTERNALLY TO THE TUBE, NOT IN CONTACT WITH THE LIQUID AND HAVE NO MOVING PARTS, THE TRANSMITTER WILL NOT BE DAMAGED BY WEAR, BY FOULING OR PRESSURE
All CONFIGURATION USER-ENTERED VALUES are saved inEE PROM, PW protected
The Transmitter is EQUIPPED WITH A CLOCK TO STORE THE MEASURE DETECTION DATE AND TIME operated by battery
IN CASE OF POWER FAILURE will be necessary TO RE-SET THE LOST. IN CASE OF INCORRECT SETTING OF THE TOTALIZER, THE OTHER FUNCTIONS ARE NOT COMPROMISED
Is composed by a DIGITAL CONVERTER AND 2 CLAMP-ON OR INSERTION ULTRASONIC TRANSDUCERS

TECHNICAL FEATURES

DN PIPE DN20 ÷ DN4.000
PIPE MATERIAL Steel, Stainless Steel, Cast Iron, Copper, PVC, Aluminium, Fiberglass (Cement with insertion trasnducers)
TOTALIZER 7 digit for differential direct and reverse flow
SPEED RANGE ±32 m/s
LINEARITY 0,5% repeatability - 0,2% total accuracy ±1%
RESPONSE TIME 1÷999 sec. programmable
DISPLAY 2x2016 alphanumeric characters
KEYPAD 4 membrane keys
DATA SHOWN instant flow rate, total flow, other
TOTALIZERS INTERNAL VOLUME 7 digit totalizer, direct and reverse flow
SECURITY setting and changing password protected
DATA LOGGER automatic memory: total flow of the last 64 days, 64 mounths, 5 years
INPUT up to five 4÷20 mA inputs
OUTPUT selectable 4÷20 mA / 0÷20 mA
FREQUENCY OUTPUT programmable 10÷9999 Hz
OUTPUT RELAY pulse or alarm totalizer
MAX HUMIDITY 85% RH not condensing (40°C) process temperature sensor 0÷150°C
SENSOR HUMIDITY 98% RH not condensing (40°C)
POWER SUPPLY 230 VAC / 24 VDC
CASING aluminum, wall mounting